PULSE 6 SERVICE MANUAL
PAGE 7.7
2009SUNRISE MEDICAL
Section 7
Shock Removal
The shocks used in this power base are pre-
adjusted at the factory. Contact Sunrise Medical
Technical Service if adjustment is required.
1. Remove the batteries. Refer to the Battery
Removal section of this manual.
2. Remove the motor/gearbox assembly from
the side on which you are removing the shock.
Refer to Motor/Gearbox Removal section of this
manual.
3. Use a 17 mm wrench and remove the top
bolt securing the shock to the base ( g. 7.20).
4. Use two 19 mm wrenches and remove
the bottom nut and bolt securing the shock to the
base ( g. 7.21 & 7.22).
5. Lift the shock from the base.
6. Reverse the previous procedures to install
the shock.
Removal Procedures (cont.)
fig. 7.20
fig. 7.21
fig. 7.22