PULSE 6 SERVICE MANUAL
PAGE 7.3
SEPT 2011SUNRISE MEDICAL
Section 7
7. Locate the two batteries mounted in the base
(fi gure 7.3.1) Note the battery connectors.
8. Disconnect the battery connectors
(fi gure 7.3.2).
8. Remove the batteries.
9. Replace the batteries by reversing the
previous procedures.
Removal Procedures (cont.)
fig. 7.3.1
fig. 7.3.2
Batteries installed incorrectly can blow the fuse
for this system. Pay careful attention to install the
battery harness across both batteries instead of to
each of the batteries individually.
Battery Installation