PULSE 6 SERVICE MANUAL
PAGE 1.2
SEPT 2011 SUNRISE MEDICAL
Troubleshooting: No Power (cont.)
Check Battery Wire Harness
Check that the battery wire harness has the
correct polarity. Set the meter to dc volts and
measure the connector with the red lead on the
+ terminal and the black lead on the negative ter-
minal as shown in (figure 1.2.1). If the voltage is
absent proceed to battery fuse test. If the polarity
is reversed correct battery wiring.
Fig 1.2.1
Check that the battery fuse is in good condition.
With the batteries disconnected, set the meter
to ohms and measure the resistance across the
fuse. see (figure 1.2.2). If the meter reads more
than one ohm, change the wiring harness, or else
proceed to the next step.
Battery Fuse
Fig 1.2.2
If the above steps did not correct the problem,
change the main harness.
Main Harness
Section 1
Circuit Breaker Test
For chairs prior to PLS-100808
To check the circuit breaker set the meter to ohms
and measure the resistance across the circuit
breaker as shown in (figure 1.2.3) if the meter
reads more than 1 ohm, then change the circuit
breaker, otherwise proceed to next step.
Fig 1.2.3